FDA Device recall Z-0440-2022
Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human ser…
- FDA Device
- Class II
- terminated
- Published 2022-01-05
On 2022-01-05 the FDA classified a Class II recall of Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human ser… by Siemens Healthcare Diagnostics, citing ¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approxim….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0440-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-12-01 |
| Published | 2022-01-05 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
Atellica CH ¿2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma Siemens Material Number (SMN): 11097635
Reason
¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L
Identifiers
| code_info | Lot Number: 211446 UDI: 00630414595566 |
|---|