QUADROX-i Neonatal with integrated arterial filter and VHK 11000 (VKMO 11000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055207 The membr…

On 2018-12-05 the FDA classified a Class I recall of QUADROX-i Neonatal with integrated arterial filter and VHK 11000 (VKMO 11000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055207 The membr… by Maquet Cardiovascular Us Sales, citing there is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before pa….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0437-2019
ClassificationClass I
Statusterminated
Initiated2018-10-25
Published2018-12-05
Last updated2026-09-13 11:54 UTC
CompanyMaquet Cardiovascular Us Sales
DistributionUS

Product

QUADROX-i Neonatal with integrated arterial filter and VHK 11000 (VKMO 11000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055207 The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.

Reason

There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.

Identifiers

code_infoUDI: 04037691773414; Lots: 92209374, 92211752, 92215106, 92222131, 92223169, 92228805, 92234286, 92236876