NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee pros…
- FDA Device
- Class II
- terminated
- Published 2012-12-05
On 2012-12-05 the FDA classified a Class II recall of NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee pros… by Zimmer, citing zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation tha….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0437-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-09-21 |
| Published | 2012-12-05 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Zimmer |
| Distribution | AZ, CA, FL, IA, IL, IN, MO, NE, NJ, NY, OH, PA, WA |
Product
NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee prostheses. The device is indicated for patients with severe knee pain and disability due to rheumatoid arthritis; osteoarthritis; traumatic arthritis; polyarthritis; collagen disorders and/or avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (particularly when there is patellofemoral erosion, dysfunction or prior patellectomy); moderate valgus, varus or flexion deformities; salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
Reason
Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated parameters. As a result, the devices may contain residual particulate from the manufacturing process. Risks include: Allergic reaction due to the foreign material or late onset
Identifiers
| code_info | Catalog 00-5986-047-02, lots 62049114 and 62045235. |
|---|