RecallRadar

FDA Device recall Z-0437-2013

NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee pros…

On 2012-12-05 the FDA classified a Class II recall of NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee pros… by Zimmer, citing zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation tha….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0437-2013
ClassificationClass II
Statusterminated
Initiated2012-09-21
Published2012-12-05
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionAZ, CA, FL, IA, IL, IN, MO, NE, NJ, NY, OH, PA, WA

Product

NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee prostheses. The device is indicated for patients with severe knee pain and disability due to rheumatoid arthritis; osteoarthritis; traumatic arthritis; polyarthritis; collagen disorders and/or avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (particularly when there is patellofemoral erosion, dysfunction or prior patellectomy); moderate valgus, varus or flexion deformities; salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.

Reason

Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated parameters. As a result, the devices may contain residual particulate from the manufacturing process. Risks include: Allergic reaction due to the foreign material or late onset

Identifiers

code_infoCatalog 00-5986-047-02, lots 62049114 and 62045235.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0437-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.