RecallRadar

FDA Device recall Z-0436-2019

QUADROX-i Neonatal without integrated arterial filter and VHK 11000 (VKMO 10000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055120 The me…

On 2018-12-05 the FDA classified a Class I recall of QUADROX-i Neonatal without integrated arterial filter and VHK 11000 (VKMO 10000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055120 The me… by Maquet Cardiovascular Us Sales, citing there is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before pa….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0436-2019
ClassificationClass I
Statusterminated
Initiated2018-10-25
Published2018-12-05
Last updated2026-09-13 11:54 UTC
CompanyMaquet Cardiovascular Us Sales
DistributionUS

Product

QUADROX-i Neonatal without integrated arterial filter and VHK 11000 (VKMO 10000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055120 The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.

Reason

There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.

Identifiers

code_infoUDI: 04037691780139; Lots: 92212343, 92214685, 92225187

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0436-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.