FDA Device recall Z-0436-2016

Optional PS500 Power Supply Unit for the Evita V500 Ventilator and Babylog VN500 Ventilator

On 2015-12-30 the FDA classified a Class I recall of Optional PS500 Power Supply Unit for the Evita V500 Ventilator and Babylog VN500 Ventilator by Draeger Medical, citing the firm became aware of cases in which the battery run times of the optional PS 500 power supply unit with the Infinity Workstation Critical Care (Evita Infinity V 500) were unex….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0436-2016
ClassificationClass I
Statusterminated
Initiated2015-12-01
Published2015-12-30
Last updated2026-09-13 11:53 UTC
CompanyDraeger Medical
DistributionUS

Product

Optional PS500 Power Supply Unit for the Evita V500 Ventilator and Babylog VN500 Ventilator. Babylog VN500 is a ventilation unit intended for the ventilation of neonatal and pediatric patients.

Reason

The firm became aware of cases in which the battery run times of the optional PS 500 power supply unit with the Infinity Workstation Critical Care (Evita Infinity V 500) were unexpectedly short due to the design of the charging algorithm in the current software. Devices used for patient transport will be a priority.

Identifiers

code_infoCatalog numbers: 8416400/8417400 distributed June 2011 October 2015.