100/120V AC, Advanced Perfusion System Platform (APS) The Terumo¿ Advanced Perfusion System 1 is indicated for use for up to 6 hours on the…

On 2012-12-05 the FDA classified a Class II recall of 100/120V AC, Advanced Perfusion System Platform (APS) The Terumo¿ Advanced Perfusion System 1 is indicated for use for up to 6 hours on the… by Terumo Cardiovascular Systems, citing terumo Cardiovascular System (TCVS) has received reports of a situation where users experienced a total loss of functionality for some System 1 units. The reports indicate that th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0435-2013
ClassificationClass II
Statusterminated
Initiated2012-11-14
Published2012-12-05
Last updated2026-09-13 11:53 UTC
CompanyTerumo Cardiovascular Systems
DistributionUS

Product

100/120V AC, Advanced Perfusion System Platform (APS) The Terumo¿ Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Reason

Terumo Cardiovascular System (TCVS) has received reports of a situation where users experienced a total loss of functionality for some System 1 units. The reports indicate that the units went blank and shut down with no sign of power and battery backup did not initiate. The result is all pumps stop, with no safety system functionality, and the battery would not be activated. The user would be limi

Identifiers

code_infoCatalog number: 801763 and all serial numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.