FDA Device recall Z-0434-2016
UniCel DxH 600 Coulter Cellular Analysis System, Software Version 1.1.1.0, Part No. B23858 The DxH Slidemaker Stainer is a fully automated…
- FDA Device
- Class II
- terminated
- Published 2015-12-23
On 2015-12-23 the FDA classified a Class II recall of UniCel DxH 600 Coulter Cellular Analysis System, Software Version 1.1.1.0, Part No. B23858 The DxH Slidemaker Stainer is a fully automated… by Beckman Coulter, citing beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creat….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0434-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-12-01 |
| Published | 2015-12-23 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Beckman Coulter |
| Distribution | US |
Product
UniCel DxH 600 Coulter Cellular Analysis System, Software Version 1.1.1.0, Part No. B23858 The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.
Reason
Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen identification (ID) but different Patient identification when manually editing pending orders at the System Manager.
Identifiers
| code_info | all serial numbers Part No. B23858 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0434-2016%22