RecallRadar

FDA Device recall Z-0432-2016

UniCel DxH 800 Coulter Cellular Analysis System, Software Version 3.0.2.0, Part No. 629029, B24465, B24802 The UniCel DxH 800/DxH 600 analy…

On 2015-12-23 the FDA classified a Class II recall of UniCel DxH 800 Coulter Cellular Analysis System, Software Version 3.0.2.0, Part No. 629029, B24465, B24802 The UniCel DxH 800/DxH 600 analy… by Beckman Coulter, citing beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creat….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0432-2016
ClassificationClass II
Statusterminated
Initiated2015-12-01
Published2015-12-23
Last updated2026-09-13 11:53 UTC
CompanyBeckman Coulter
DistributionUS

Product

UniCel DxH 800 Coulter Cellular Analysis System, Software Version 3.0.2.0, Part No. 629029, B24465, B24802 The UniCel DxH 800/DxH 600 analyzers are quantitative, multi-parameter, automated hematology analyzers for in vitro diagnostic use in screening patient populations found in clinical laboratories. The DxH 800 is available as an analyzer for use on a benchtop or with a floor stand. The DxH 600 is available as an analyzer for use on a benchtop only.

Reason

Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen identification (ID) but different Patient identification when manually editing pending orders at the System Manager.

Identifiers

code_infoall serial numbers Part Nos. 629029, B24465, B24802

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.