RecallRadar

FDA Device recall Z-0425-2023

The FAST System and the FAST PBC Prep Cartridge together are a closed, rapid and automated sample preparation system for isolating bacteria…

On 2022-12-14 the FDA classified a Class II recall of The FAST System and the FAST PBC Prep Cartridge together are a closed, rapid and automated sample preparation system for isolating bacteria… by Qvella, citing device exceeds the limitations of a Class I 510(k)-exempt device so the user manual, instructions for use, and promotional materials are being corrected to remove all references t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0425-2023
ClassificationClass II
Statusongoing
Initiated2022-11-02
Published2022-12-14
Last updated2026-09-13 11:56 UTC
CompanyQvella
DistributionAL, CA, NY, UT

Product

The FAST System and the FAST PBC Prep Cartridge together are a closed, rapid and automated sample preparation system for isolating bacterial cells directly from a positive blood culture (PBC). The cartridge and instrument isolate and concentrate bacterial cells directly from PBC by removing blood, cellular and culture media matrix components from the sample, thereby reducing the sample volume considerably while maintaining bacterial cell viability. The resulting material is a purified microbial suspension or Liquid Colony for use in downstream in vitro diagnostics.

Reason

Device exceeds the limitations of a Class I 510(k)-exempt device so the user manual, instructions for use, and promotional materials are being corrected to remove all references to downstream applications.

Identifiers

code_infoTSP-UM-001, Revisions: 2 and 3; MKT-0025, Revisions: 2 and 3; TSP-PI-001, Revisions: 2 and 3

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0425-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.