FDA Device recall Z-0424-2013
KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case Size 5 / Adult ClearSeal King LAD¿ Patient Size: >70kg P…
- FDA Device
- Class II
- terminated
- Published 2012-11-28
On 2012-11-28 the FDA classified a Class II recall of KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case Size 5 / Adult ClearSeal King LAD¿ Patient Size: >70kg P… by King Systems, citing the firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the exp….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0424-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-11-02 |
| Published | 2012-11-28 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | King Systems |
| Distribution | US |
Product
KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case Size 5 / Adult ClearSeal King LAD¿ Patient Size: >70kg Product Usage: Laryngeal airway
Reason
The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO
Identifiers
| code_info | REF: LAD-705, Lot #K120413 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0424-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.