RecallRadar

FDA Device recall Z-0423-2023

Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Chargin…

On 2022-12-14 the FDA classified a Class II recall of Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Chargin… by Insulet, citing the firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat genera….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0423-2023
ClassificationClass II
Statusongoing
Initiated2022-11-14
Published2022-12-14
Last updated2026-09-13 11:56 UTC
CompanyInsulet
DistributionUS

Product

Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM

Reason

The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.

Identifiers

code_infoALL serial numbers/ controller lots are currently in scope of this recall

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.