RecallRadar

FDA Device recall Z-0420-2026

TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01

On 2025-11-12 the FDA classified a Class II recall of TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01 by Integra Lifesciences Neurosciences, citing it was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system ma….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0420-2026
ClassificationClass II
Statusongoing
Initiated2025-10-03
Published2025-11-12
Last updated2026-09-13 11:57 UTC
CompanyIntegra Lifesciences Neurosciences
DistributionUS

Product

TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01;

Reason

It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.

Identifiers

code_infoManufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01; UDI-DI: 10846835021103;

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0420-2026%22