RecallRadar

FDA Device recall Z-0420-2014

SIGMA¿ LCS¿ HIGH PERFORMANCE" MBT PUNCH CEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal…

On 2013-12-04 the FDA classified a Class II recall of SIGMA¿ LCS¿ HIGH PERFORMANCE" MBT PUNCH CEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal… by Depuy Orthopaedics, citing dePuy Orthopaedics, Inc. is issuing a voluntary device correction of the HP M.B.T. Keel Punch Knee Instrument because of the potential for tabs on the connection end to fracture a….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0420-2014
ClassificationClass II
Statusterminated
Initiated2013-11-18
Published2013-12-04
Last updated2026-09-13 11:53 UTC
CompanyDepuy Orthopaedics
DistributionUS

Product

SIGMA¿ LCS¿ HIGH PERFORMANCE" MBT PUNCH CEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive tibial tray used during total knee arthroplasty.

Reason

DePuy Orthopaedics, Inc. is issuing a voluntary device correction of the HP M.B.T. Keel Punch Knee Instrument because of the potential for tabs on the connection end to fracture and be left in the patient.

Identifiers

code_infoItem #950502018

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0420-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.