FDA Device recall Z-0417-2022
Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument Kit, Mode: 9736188, and the Bundl…
- FDA Device
- Class II
- ongoing
- Published 2021-12-29
On 2021-12-29 the FDA classified a Class II recall of Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument Kit, Mode: 9736188, and the Bundl… by Medtronic Navigation, citing A tracker that holds and guides neurosurgical surgical instruments may be subject to separation/weld failure, where the tracker connects to the instrument tube. If a weld separati….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0417-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-11-09 |
| Published | 2021-12-29 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Medtronic Navigation |
| Distribution | AZ, CA, FL, IL, IN, KS, KY, MI, MN, MO, MS, NE, NV, OH, OK, OR, SC, TX, UT, VA, WA |
Product
Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument Kit, Mode: 9736188, and the Bundle Autoguide Reg Kit.
Reason
A tracker that holds and guides neurosurgical surgical instruments may be subject to separation/weld failure, where the tracker connects to the instrument tube. If a weld separation occurs and a navigation inaccuracy is experienced, it may result in prolonged procedure and tissue injury. Per 11/24/21 firm email: The Bundle Autoguide Reg Kit does not have a model number. The Bundle Autoguide Reg Kit is internal terminology and not visible to the customer and are used as sales and shipping configuration only. When this sales bundle is chosen by an internal Medtronic Sales Representative the devices included in the bundle are individually packaged and labeled and sent to the customer in a single shipment. The devices included in the bundle are ordered, packed, and shipped as individual medical devices with unique model numbers, which is the reason that the Autoguide Tracker (28248) and the Stealth Autoguide Basic Instrument Kit (9736188) are the only part numbers subject to this FCA.
Identifiers
| code_info | All lot numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0417-2022%22