RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary complications as a result of immobility, Thoracic…

On 2012-11-28 the FDA classified a Class II recall of RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary complications as a result of immobility, Thoracic… by Kci Usa, citing the firm is initiating a voluntary medical device correction for the RotoRest Delta Therapy System head pack. KCI has received complaints alleging patients experiencing skin break….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0409-2013
ClassificationClass II
Statusterminated
Initiated2012-06-27
Published2012-11-28
Last updated2026-09-13 11:53 UTC
CompanyKci Usa
DistributionUS

Product

RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary complications as a result of immobility, Thoracic or lumbar fracture, Cervical traction, Skeletal traction

Reason

The firm is initiating a voluntary medical device correction for the RotoRest Delta Therapy System head pack. KCI has received complaints alleging patients experiencing skin breakdown on the head and face due to the positioning of the head pack. In a few instances, the patient received injuries ranging from Stage III to Stage IV pressure ulcers.

Identifiers

code_infoModel Number 208030, Serial Numbers: 00157087, 00157091, 00157288, 00157375, 00157405, RRCKOO011, RRHKOO065, RRHKOO074, RRHKOO096, RRHK00109, and RRHKOO117.