FDA Device recall Z-0403-2016
Olympus ENDOEYE HD II Video Telescopes
- FDA Device
- Class II
- terminated
- Published 2015-12-16
On 2015-12-16 the FDA classified a Class II recall of Olympus ENDOEYE HD II Video Telescopes by Olympus, citing A damaged temperature sensor at the ENDOEYE tip caused the distal end to become abnormally hot. Excessive heating of the ENDOEYE distal end could result in patient or user injury.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0403-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-11-17 |
| Published | 2015-12-16 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Olympus |
| Distribution | US |
Product
Olympus ENDOEYE HD II Video Telescopes. Models WA50040A ENDOEYE HD II, 10 mm, 0 degree, WA50042A ENDOEYE HD II 10 mm, 30 degree, W7100603 Repair cable 10 mm 0/30 degree. The ENDOEYE video Telescopes are used with other supporting equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.
Reason
A damaged temperature sensor at the ENDOEYE tip caused the distal end to become abnormally hot. Excessive heating of the ENDOEYE distal end could result in patient or user injury.
Identifiers
| code_info | Serial numbers 622483, 622523, 622553, 622566, 622568, 622569, 622573, 622576, 622614, 622615, 622616, 622617, 622618,…Serial numbers 622483, 622523, 622553, 622566, 622568, 622569, 622573, 622576, 622614, 622615, 622616, 622617, 622618, 622619, 622620, 622632, 622643, 622644, 622645, 622646, 622647, 622654, 622655, 622656, 622657, 622658, 622664, 622666, 622667, 622687, 622688, 622693, 622704, 622705, 622706, 622710, 622711, 622712, 622718, 622719, 622720, 622722, 622723, 622750, 622753, 622764, 622766, 622767, 622769, 622771, 622773, 622775, 622777, 622778, 622785, 622789, 622795, 622797, 622798, 622801, 622806, 623624, 623626, 623629, 623637, 623638, 623639, 623641, 623661, 623669, 623671, 623672, 623673, 623677, 623687, 623688, 623705, 623709, 623713, 623719, 623720, 623722, 623723, 623729, 623730, 623731, 623735, 623736, 623737, 623738, 623739, 623740, 623741, 623742, 623743, 623747, 623749, 623750, 623751, 623752, 623776, 623778, 623784, 623786, 623787, 623788, 623799, 623800, 623801, 623802, 624004, 624005, 624006, 624008, 624011, 624012, 624016, 624017, 624018, 624019, 624024, 624028, 624030, 6 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0403-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.