FDA Device recall Z-0396-2021

ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 59 mm Number: 00-8065-956-32

On 2020-11-11 the FDA classified a Class II recall of ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 59 mm Number: 00-8065-956-32 by Zimmer Biomet, citing sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the pote….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0396-2021
ClassificationClass II
Statusterminated
Initiated2020-09-24
Published2020-11-11
Last updated2026-09-13 11:55 UTC
CompanyZimmer Biomet
DistributionUS

Product

ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 59 mm Number: 00-8065-956-32

Reason

Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.

Identifiers

code_infoAll lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.