RecallRadar

FDA Device recall Z-0393-2013

. 00789503601 VerSys Femoral Head Provisional 36mm -3.5 00789503602 VerSys Femoral Head Provisional 36mm +0 00789503603 VerSys Femoral Head…

On 2012-11-28 the FDA classified a Class II recall of . 00789503601 VerSys Femoral Head Provisional 36mm -3.5 00789503602 VerSys Femoral Head Provisional 36mm +0 00789503603 VerSys Femoral Head… by Zimmer, citing surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0393-2013
ClassificationClass II
Statusterminated
Initiated2010-10-10
Published2012-11-28
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

. 00789503601 VerSys Femoral Head Provisional 36mm -3.5 00789503602 VerSys Femoral Head Provisional 36mm +0 00789503603 VerSys Femoral Head Provisional 36mm +3.5 00789503604 VerSys Femoral Head Provisional 36mm +7 00789503605 VerSys Femoral Head Provisional 36mm +10.5 12/14 Neck Taper The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.

Reason

Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat

Identifiers

code_infoAll lots distributed from lot 69998600 (2/20/1996) to present

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0393-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.