FDA Device recall Z-0389-2013

Product 100789502202 VerSys Femoral Head Provisional 22mm +0 00789502203 VerSys Femoral Head Provisional 22mm +3.5 00789502220 VerSys Femor…

On 2012-11-28 the FDA classified a Class II recall of Product 100789502202 VerSys Femoral Head Provisional 22mm +0 00789502203 VerSys Femoral Head Provisional 22mm +3.5 00789502220 VerSys Femor… by Zimmer, citing surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0389-2013
ClassificationClass II
Statusterminated
Initiated2010-10-10
Published2012-11-28
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

Product 100789502202 VerSys Femoral Head Provisional 22mm +0 00789502203 VerSys Femoral Head Provisional 22mm +3.5 00789502220 VerSys Femoral Head Provisional 22mm -2 00789502230 VerSys Femoral Head Provisional 22mm +3 12/14 Neck Taper The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.

Reason

Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat

Identifiers

code_infoAll lots distributed from lot 69998600 (2/20/1996) to present