FDA Device recall Z-0388-2018
Brilliance iCT Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different ang…
- FDA Device
- Class II
- terminated
- Published 2018-01-24
On 2018-01-24 the FDA classified a Class II recall of Brilliance iCT Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different ang… by Philips Medical Systems Cleveland, citing philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occ….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0388-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-11-10 |
| Published | 2018-01-24 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Philips Medical Systems Cleveland |
| Distribution | n/a |
Product
Brilliance iCT Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason
Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.
Identifiers
| code_info | Code 728306, SN# 100776, UDI: (01)00884838059474(21)100776; 100778, UDI: (01)00884838059474(21)100778; 100779, UDI: (…Code 728306, SN# 100776, UDI: (01)00884838059474(21)100776; 100778, UDI: (01)00884838059474(21)100778; 100779, UDI: (01)00884838059474(21)100779; 100780, UDI: (01)00884838059474(21)100780; 100781, UDI: (01)00884838059474(21)100781; 100782, UDI: (01)00884838059474(21)100782; 100783, UDI: (01)00884838059474(21)100783; 100784, UDI: (01)00884838059474(21)100784; 100785, UDI: (01)00884838059474(21)100785; 100787, UDI: (01)00884838059474(21)100787; 100790, UDI: (01)00884838059474(21)100790; 100791, UDI: (01)00884838059474(21)100791; and 100794, UDI: (01)00884838059474(21)1007. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.