FDA Device recall Z-0387-2017
Roche COBAS INTEGRA c111 Analyzer, Chemistry (Photometric, Discrete), for clinical use Product Usage: The Roche COBAS INTEGRA c111 analyzer…
- FDA Device
- Class II
- terminated
- Published 2016-11-16
On 2016-11-16 the FDA classified a Class II recall of Roche COBAS INTEGRA c111 Analyzer, Chemistry (Photometric, Discrete), for clinical use Product Usage: The Roche COBAS INTEGRA c111 analyzer… by Roche Diagnostics Operations, citing cobas c 111 analyzers (catalog numbers 04777433001 and 04528778001) with software versions up to and including 4.20 may encounter the following alarm: 7002: 108000572, A software….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0387-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-09-27 |
| Published | 2016-11-16 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Roche Diagnostics Operations |
| Distribution | US |
Product
Roche COBAS INTEGRA c111 Analyzer, Chemistry (Photometric, Discrete), for clinical use Product Usage: The Roche COBAS INTEGRA c111 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically. The analyzer also has an optional ISE module for measuring sodium, potassium and chloride.
Reason
cobas c 111 analyzers (catalog numbers 04777433001 and 04528778001) with software versions up to and including 4.20 may encounter the following alarm: 7002: 108000572, A software error occurred. This alarm is generated due to a measurement timing error. Under very rare conditions, the instrument may process two tests in the same cuvette if the run restarts. If a used cuvette is used again result of the test(s) will be erroneous. These erroneous results may not be flagged. Falsely low or high patient results may lead to incorrect diagnostic measures and medical therapeutic decisions. The medical risk depends on the parameter.
Identifiers
| code_info | 04777433001 Cobas c111 with ISE 04528778001 Cobas c111 without ISE cobas c 111 analyzers (catalog numbers 04777…04777433001 Cobas c111 with ISE 04528778001 Cobas c111 without ISE cobas c 111 analyzers (catalog numbers 04777433001 and 04528778001) with software versions up to and including 4.20 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0387-2017%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.