RecallRadar

FDA Device recall Z-0387-2014

Large Buttress Compression Nut For Blade Guide Sleeve TFN The Synthes Large Buttress Compression Nut For Blade Guide Sleeve TFN is intended…

On 2013-12-04 the FDA classified a Class II recall of Large Buttress Compression Nut For Blade Guide Sleeve TFN The Synthes Large Buttress Compression Nut For Blade Guide Sleeve TFN is intended… by Synthes Usa Hq, citing the Large Buttress Compression Nut For Blade Guide Sleeve TFN was recalled because of complaints that the compression nut did not thread completely onto the blade guide sleeve and….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0387-2014
ClassificationClass II
Statusterminated
Initiated2013-07-31
Published2013-12-04
Last updated2026-09-13 11:53 UTC
CompanySynthes Usa Hq
DistributionAR, AZ, CA, CO, GA, IL, IN, MI, MS, NY, PA, TN, TX, VA

Product

Large Buttress Compression Nut For Blade Guide Sleeve TFN The Synthes Large Buttress Compression Nut For Blade Guide Sleeve TFN is intended to treat stable and unstable pertrochanteric fractures, intertrochanteric fractures, basal neck fractures, and combinations there of.

Reason

The Large Buttress Compression Nut For Blade Guide Sleeve TFN was recalled because of complaints that the compression nut did not thread completely onto the blade guide sleeve and became stuck.

Identifiers

code_infoPart # Lot # SD357.371 5734551 6083065 6538811 6631975 6703191 6703190 6841538

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0387-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.