RecallRadar

FDA Device recall Z-0381-2013

Stryker MANTIS TORQUE WRENCH REF 48287028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCE +33 (0)5.57.97.06.30 http://ww…

On 2012-11-28 the FDA classified a Class II recall of Stryker MANTIS TORQUE WRENCH REF 48287028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCE +33 (0)5.57.97.06.30 http://ww… by Stryker Spine, citing multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0381-2013
ClassificationClass II
Statusterminated
Initiated2012-08-09
Published2012-11-28
Last updated2026-09-13 11:53 UTC
CompanyStryker Spine
DistributionUS

Product

Stryker MANTIS TORQUE WRENCH REF 48287028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCE +33 (0)5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine, 2 Pearl Ct. Allendale NJ 07401-1677 USA +1-201-780-8000 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.

Reason

Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

Identifiers

code_infoClass I device, 510(K) exempt Catalog Number: 48287028 Lot Number: 098400

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0381-2013%22