Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal

On 2018-02-07 the FDA classified a Class II recall of Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal by Carl Zeiss Metrology, citing due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray gener….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0379-2018
ClassificationClass II
Statusongoing
Initiated2017-11-14
Published2018-02-07
Last updated2026-09-13 11:54 UTC
CompanyCarl Zeiss Metrology
DistributionAZ, CA, FL, GA, IL, NC, NJ, PR, WI

Product

Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal.

Reason

Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.

Identifiers

code_infoSerial Numbers: 152806 157499 157683 181193 181364 182887 186654 186767 186797 192099 192182 192351 197342