ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma

On 2016-11-16 the FDA classified a Class II recall of ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma by Abbott Laboratories, citing abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21).

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0379-2017
ClassificationClass II
Statusterminated
Initiated2016-10-11
Published2016-11-16
Last updated2026-09-13 11:54 UTC
CompanyAbbott Laboratories
DistributionUS

Product

ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma.

Reason

Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21).

Identifiers

code_info
Lot: 31497UN14 Exp.: 17 Oct. 2016; Lot: 09353UNI15 Exp.: 12 Feb. 2017; Lot: 45216UN15 12 May 2017; Lot: 14596UN15 Exp.:…Lot: 31497UN14 Exp.: 17 Oct. 2016; Lot: 09353UNI15 Exp.: 12 Feb. 2017; Lot: 45216UN15 12 May 2017; Lot: 14596UN15 Exp.: 30 Jul. 2017; Lot: 37055UN15 Exp.: 12 Oct. 2017; Lot: 14782UN16 Exp.: 21 Apr. 2018