FDA Device recall Z-0377-2026

Model: Incisive CT for Brazil SKD

On 2025-11-05 the FDA classified a Class II recall of Model: Incisive CT for Brazil SKD by Philips Medical Systems, citing issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0377-2026
ClassificationClass II
Statusongoing
Initiated2025-09-25
Published2025-11-05
Last updated2026-09-13 11:57 UTC
CompanyPhilips Medical Systems
DistributionAL, AR, CA, CT, FL, GA, IL, IN, KY, MD, MI, OH, PA, PR, TN, TX, UT, VA

Product

Model: Incisive CT for Brazil SKD; Product Code (REF): 728146; Software Versions: 5.1.0.X & 5.1.1.X;

Reason

Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.

Identifiers

code_infoProduct Code (REF): 728146; UDI-DI: None; Serial Number: 540000

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.