Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile (The TFN-ADVANCED Proximal Femoral Nailing System…

On 2015-12-16 the FDA classified a Class II recall of Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile (The TFN-ADVANCED Proximal Femoral Nailing System… by Synthes Usa Products, citing TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulated connecting screw from fully tightening the insertion handle to the n….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0375-2016
ClassificationClass II
Statusterminated
Initiated2015-11-02
Published2015-12-16
Last updated2026-09-13 11:53 UTC
CompanySynthes Usa Products
DistributionCA, FL, MI, WI

Product

Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile (The TFN-ADVANCED Proximal Femoral Nailing System) The TFN-ADVANCED Proximal Femoral Nailing System is intended for treatment of fractures. Intramedullary fixation rod, orthopedic.

Reason

TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulated connecting screw from fully tightening the insertion handle to the nail resulting in a loose or toggling nail in the handle. Or 2. Cause the bottom of the cannulated connecting screw to tighten against the top of the locking mechanism preventing advancement of the locking mechanism.

Identifiers

code_infoPart Number Lot Number 04.037.162S 7950511