RecallRadar

FDA Device recall Z-0373-2014

GE Vivid E9 ultrasound system

On 2013-11-27 the FDA classified a Class II recall of GE Vivid E9 ultrasound system by Ge Healthcare, citing GE became aware of a potential safety issue due to a system slow down and system lock up associated with the software of Vivid E9 ultrasound system. During a critical procedure th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0373-2014
ClassificationClass II
Statusterminated
Initiated2013-10-11
Published2013-11-27
Last updated2026-09-13 11:53 UTC
CompanyGe Healthcare
DistributionUS

Product

GE Vivid E9 ultrasound system. GE Vivid E9 is a Track 3 diagnostic ultrasound system, which is primarily intended for cardiac imaging and analysis, but which also includes vascular and general radiology applications. The Vivid E9 incorporates a variety of electronic array transducers operating in linear, curved linear, sector/phased array or matrix array format, including two dedicated CW transducers and several real time 3D transducers. It consists of a mobile console with keyboard control panel; color LCD/TFT touch panel, LCD color video display and optional image storage and printing devices. It provides high performance ultrasound imaging and analysis and has comprehensive networking and DICOM capability.

Reason

GE became aware of a potential safety issue due to a system slow down and system lock up associated with the software of Vivid E9 ultrasound system. During a critical procedure the Vivid E9 Scanner may gradually become unresponsive and eventually lock up, with the result that the vivid E9 Scanner must be rebooted.

Identifiers

code_info
Serial numbers: 0000000VE94166, 0000000VE94167, 0000000VE94170, 0000000VE94171, 0000000VE94172, 0000000VE94173, 00…Serial numbers: 0000000VE94166, 0000000VE94167, 0000000VE94170, 0000000VE94171, 0000000VE94172, 0000000VE94173, 0000000VE94174, 0000000VE94175, 0000000VE94176, 0000000VE94177, 0000000VE94178, 0000000VE94179, 0000000VE94180, 0000000VE94181, 0000000VE94182, 0000000VE94183, 0000000VE94184, 0000000VE94185, 0000000VE94186, 0000000VE94187, 0000000VE94188, 0000000VE94190, 0000000VE94191, 0000000VE94192, 0000000VE94193, 0000000VE94194, 0000000VE94195, 0000000VE94196, 0000000VE94197, 0000000VE94198, 0000000VE94199, 0000000VE94210, 0000000VE94211, 0000000VE94212, 0000000VE94213, 0000000VE94214, 0000000VE94215, 0000000VE94216, 0000000VE94217, 0000000VE94225, 0000000VE94226, 0000000VE94227, 0000000VE94231, 0000000VE94232, 0000000VE94233, 0000000VE94234, 0000000VE94235, 0000000VE94236, 0000000VE94237, 0000000VE94238, 0000000VE94243, 0000000VE94244, 0000000VE94245, 0000000VE94246, 0000000VE94247, 0000000VE94248, 0000000VE94249, 0000000VE94250, 0000000VE94251, 0000000VE94252, 0000000VE94253, 000

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0373-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.