RecallRadar

FDA Device recall Z-0369-2017

TBS iNsight Version v.3.0.1 Product Usage: TBS iNsight is a medical device software that is installed on bone densitometers for analysis of…

On 2016-11-16 the FDA classified a Class II recall of TBS iNsight Version v.3.0.1 Product Usage: TBS iNsight is a medical device software that is installed on bone densitometers for analysis of… by Medimaps Group, citing the FRAX adjusted for TBS values are not correct when: The FRAX feature is activated in TBS iNsight; and TBS has been computed from a spine scan where some vertebrae were excluded.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0369-2017
ClassificationClass II
Statusterminated
Initiated2016-10-03
Published2016-11-16
Last updated2026-09-13 11:54 UTC
CompanyMedimaps Group
DistributionUS

Product

TBS iNsight Version v.3.0.1 Product Usage: TBS iNsight is a medical device software that is installed on bone densitometers for analysis of bone microarchitecture and osteoporosis management.

Reason

The FRAX adjusted for TBS values are not correct when: The FRAX feature is activated in TBS iNsight; and TBS has been computed from a spine scan where some vertebrae were excluded.

Identifiers

code_infoVersion v.3.0.1; Serial # 14295, 300409, 70580, 200042, 500581, 70103, 70174, 80936, 100202, 201120, 84584, 83769, 100112, 210020, 210440

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0369-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.