RecallRadar

FDA Device recall Z-0367-2025

Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications includi…

On 2024-11-20 the FDA classified a Class II recall of Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications includi… by Stryker, citing there is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0367-2025
ClassificationClass II
Statusongoing
Initiated2024-10-04
Published2024-11-20
Last updated2026-09-13 11:57 UTC
CompanyStryker
DistributionUS

Product

Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-001

Reason

There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries

Identifiers

code_infoGTIN: 07613327117219 Lot Number: 64117291

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0367-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.