FDA Device recall Z-0367-2014
GE Healthcare Discovery NMCT 670, Discovery NM 630, Optima NMCT 640, & Brivo NM 615
- FDA Device
- Class II
- terminated
- Published 2013-11-27
On 2013-11-27 the FDA classified a Class II recall of GE Healthcare Discovery NMCT 670, Discovery NM 630, Optima NMCT 640, & Brivo NM 615 by Ge Healthcare, citing GE Healthcare has recently become aware of a potential safety issue related to a SPECT scan procedure used for pulmonary imaging. Contact between the Detector (GE Healthcare Disco….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0367-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-10-25 |
| Published | 2013-11-27 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ge Healthcare |
| Distribution | US |
Product
GE Healthcare Discovery NMCT 670, Discovery NM 630, Optima NMCT 640, & Brivo NM 615. The GE Discovery NMJCT 670 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce: NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body scanning) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). The acquisition types include single and multi-isotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include assortment of collimators, gating by physiological signals, and real-time automatic body contouring. CT System: Cross sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes, including Axial, Cine, Helical, Cardiac, and Gated acquisitions. These images may be obtained with or without contrast. The CT system is indicated for head, whole body, cardiac and vascular X-Ray Computed Tomography applications NM + CT System: Combined, hybrid SPECT and CT protocols, for CT-based SPECT attenuation corrected imaging as well as functional and anatomical mapping imaging (localization, registration and fusion). The GE Discovery NM/CT 670 system may include signal analysis and display equipment, patient and equipment supports, components and accessories. The system may include data and image processing to produce images in a variety of trans-axial and reformatted planes. The images can also be post processed to produce additional images, imaging planes, and analysis results. The system may be used for patients of all ages. The GE Discovery NM 630 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM 630 system is a Nuclear Medicine (NM) system, which is intended to yield General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body scanning) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). The acquisition types include single and multi isotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include assortment of collimators, gating by physiological signals, and real-time automatic body contouring. The GE Discovery NM 630 system may include signal analysis and display equipment, patient and equipment supports, components and accessories. The system may be used for patients of all ages. - The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine (NM) system using a variety of scann
Reason
GE Healthcare has recently become aware of a potential safety issue related to a SPECT scan procedure used for pulmonary imaging. Contact between the Detector (GE Healthcare Discovery NM/CT 670, and/or Optima NM/CT 640, and/or Discovery NM630, and/or Brivo NM 615 all configurations) and the Patient's elbow may occur during SPECT protocols using 450 angular steps in scenarios when a Patients a
Identifiers
| code_info | Serial # System ID 21000 262574D670 21023 309671SPECT 21030 856247NM 21032 617638DSV670 21044 404686NMCT 21045 97…Serial # System ID 21000 262574D670 21023 309671SPECT 21030 856247NM 21032 617638DSV670 21044 404686NMCT 21045 972935D670 21054 262574D630 21062 305674D670 21065 PITT670 21069 212639NMD670 21074 423495D670 21078 615322NM5 21105 803751D670 21111 DUKECCDSV670NM 21114 614NCH670 21117 262574O640 21127 410D6701 21149 540332D670 21157 856247D670 21172 630432D670 21182 GON1409671 21188 415353D670 21189 DUKEDMP670NM 21193 610402D670NM1 21201 720848D670 21202 972747NM670 21221 410955D670 21223 307362D670 21224 660890D670 22002 619474D630 22004 618443DISC1 22007 814623D630 22010 251639D630 22012 704355D630 22013 956523D630 22017 845454D630 22018 618842DISC1 22024 601354D630 22025 773282D630 22026 813558D630 22027 941LWRNM630 22029 PRINCE630 22030 281540D630 22032 417781D630 22033 601288D630 22034 913345D630 22036 318798D630 22039 614NCH630 22040 904321DSC1 22041 317887D630 22042 318448D630 22043 JHDSV630NM 22045 217544PH630 22046 301877D630 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0367-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.