FDA Device recall Z-0365-2023
FULL VISION INC. TRACKMASTER,TREADMILL TMX428 220V, Model #317-07927
- FDA Device
- Class II
- ongoing
- Published 2022-12-14
On 2022-12-14 the FDA classified a Class II recall of FULL VISION INC. TRACKMASTER,TREADMILL TMX428 220V, Model #317-07927 by Full Vision, citing this has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magne….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0365-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-08-31 |
| Published | 2022-12-14 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Full Vision |
| Distribution | GA, NY |
Product
FULL VISION INC. TRACKMASTER,TREADMILL TMX428 220V, Model #317-07927
Reason
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
Identifiers
| code_info | UDI/DI 00860176000613, Serial Numbers: FVDC-7630, FVDC-7631, FVDC-7633, FVDC-7634, FVDC-7635, FVDC-7636, FVDC-7637, FV…UDI/DI 00860176000613, Serial Numbers: FVDC-7630, FVDC-7631, FVDC-7633, FVDC-7634, FVDC-7635, FVDC-7636, FVDC-7637, FVDC-7638, FVDC-7639, FVDC-7647, FVDC-7648, FVDC-7649, FVDC-7650, FVDC-7651, FVDC-7652, FVDC-7653, FVDC-7654, FVDC-7655, FVDC-7656, FVDC-7657, FVDC-7658, FVDC-7659, FVDC-7660, FVDC-7661, FVDC-7662, FVDC-7663, FVDC-7664, FVDC-7665, FVDC-7707, FVDC-7708, FVDC-7709, FVDC-7710, FVDC-7711, FVDC-7712, FVDC-7713, FVDC-7714, FVDC-7715, FVDC-7716 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0365-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.