RecallRadar

FDA Device recall Z-0364-2025

Brand Name: UroPass Ureteral Access Sheaths Product Name: Ureteral Access Sheaths Model/Catalog Number: 61024BX, 61038BX, 61046BX, 61054BX,…

On 2024-11-20 the FDA classified a Class II recall of Brand Name: UroPass Ureteral Access Sheaths Product Name: Ureteral Access Sheaths Model/Catalog Number: 61024BX, 61038BX, 61046BX, 61054BX,… by Olympus, citing exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patient….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0364-2025
ClassificationClass II
Statusongoing
Initiated2024-09-11
Published2024-11-20
Last updated2026-09-13 11:57 UTC
CompanyOlympus
DistributionUS

Product

Brand Name: UroPass Ureteral Access Sheaths Product Name: Ureteral Access Sheaths Model/Catalog Number: 61024BX, 61038BX, 61046BX, 61054BX, 61124BX, 61138BX, 61146BX, 61154BX, 61224BX, 61238BX, 61246BX, 61254BX, 61324BX, 61338BX, 61346BX, 61354BX Software Version: N/A Product Description: The Olympus UroPass Ureteral Access Sheath Set consists of a hydrophilic coated outer sheath and an inner tapered dilator intended to establish a conduit for the passage of endoscopes and retrieval devices into the ureter. The hydrophilic coating on the UroPass Ureteral Access Sheath eases passage and placement. Both the outer sheath and inner dilator are radio-opaque for ease of viewing radiographically. Intended for single use only. Component: N/A

Reason

Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.

Identifiers

code_info
Model/UDI-DI: 61024BX/00821925035317, 61038BX/00821925035324, 61046BX/00821925035331, 61054BX/00821925035348, 61124BX/0…Model/UDI-DI: 61024BX/00821925035317, 61038BX/00821925035324, 61046BX/00821925035331, 61054BX/00821925035348, 61124BX/00821925035355, 61138BX/00821925035362, 61146BX/00821925035379, 61154BX/00821925035386, 61224BX/00821925035393, 61238BX/00821925035409, 61246BX/00821925035416, 61254BX/00821925035423, 61324BX/00821925035430, 61338BX/00821925035447, 61346BX/00821925035454, 61354BX/00821925035461.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0364-2025%22