FDA Device recall Z-0362-2020

TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 - Product Usage: The Thermo-Re…

On 2019-11-20 the FDA classified a Class II recall of TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 - Product Usage: The Thermo-Re… by Tec Com, citing the action is being initiated due to potential installation of faulty peltier-elements within the thermo-regulation system. Failure of the peltier-element would report a system fa….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0362-2020
ClassificationClass II
Statusterminated
Initiated2018-08-14
Published2019-11-20
Last updated2026-09-13 11:55 UTC
CompanyTec Com
DistributionUS

Product

TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 - Product Usage: The Thermo-Regulation System TECOTHERM NEO is designed for controlled cold & heat treatment procedures. By means of a mattress, cold and heat is provided to the patient, depending on the therapy objective.

Reason

The action is being initiated due to potential installation of faulty peltier-elements within the thermo-regulation system. Failure of the peltier-element would report a system failure and no longer be able to guarantee temperature control at full power.

Identifiers

code_infoSerial Numbers: 2018/09/07, 2018/10/01, 2018/10/02, 2018/10/03, 2018/10/05