FDA Device recall Z-0361-2024
Ventlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations to include: 1. AirFlow Adult…
- FDA Device
- Class II
- ongoing
- Published 2023-11-29
On 2023-11-29 the FDA classified a Class II recall of Ventlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations to include: 1. AirFlow Adult… by Sunmed Holdings, citing A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0361-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-10-11 |
| Published | 2023-11-29 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Sunmed Holdings |
| Distribution | US |
Product
Ventlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations to include: 1. AirFlow Adult Resuscitator REF #s: AF1140MB-M5, AF1140MB-D5, AF5142OB, AF5142MB, AF5142MB-D235, AF1142MBP, AF5142MBP, AF5140MB, AF5140MB-2, AF5140MB-D6, AF5140MB-F, AF1140MB-T, AF1140MB-PW, AF5140MB-PW, AF1140MB-I, AF5140MB-I, AF5140MBN, AF1140MBS, AF5140MBS, AF1140MB-X, AF5140MB-X, AF1140MB-P, AF5140MBP, AF1140MB-K, AF5140MBS-I,AF5140MBPF-G10, AF5140MBP-G10, AF1140MBP-T, AF5140MBP-I, AF1140MBP-K, AF1140MBC-N, AF1140MBP-2, AF1140MC, AF5140MC, AF1140MCP, AF5140MC-P, AF1100MB, AF5100MB, AF5100MC, AF1140MC-M5, AF1140MB-TM5, AF1140MBP-TM5, AF1100MC, AF1140OB, AF1140MBP-M5, AFW1140MB,AF1140MBS-P, AF1000MBN-20, AF5102MBP, AF1100MB-N, AF1100MC-P, AF1140MB, AF1140MB-E, AF1140MB-N, AF1140MB-R, AF1140MBS-K, AF1140MBSP-I, AF1142MB, AF1100MBS, AF1100MB-MMC, AF1100MB-PWG1, AF1100MBP; 2. AirFlow Child Resuscitator REF #s: AF4100MCP, AF4140MB, AF4140MB-I, AF4140MBP, AF4140MB-PI, AF4100MB, AF4100MBP, AF4100OB, AF4140MB-D23, AF4140OBP, AF4140MCP-D3, AF4140MC; 3. AirFlow Infant Resuscitator REF #s: AF3140OB, AF3144MB, AF3104MB-PW, AF3100MBP- M0D1, AF3100MBP- M012, AF3140MB-M000, AF3140MB-D2, AF3140MB-D1, AF3140MB-S1, AF3140MB-2B, AF3140MB-I, AF3140MB-K, AF3100MB-PW, AF3100MB-PTS1, AF3100MB-M0- D1, AF3100OC, AF3100MC-2, AF3100MC, AF3106MC-2, AF3140MCP-K, AF3140MC-2B, AF3100MCP-2, AF3140MC- M0D12, AF3144MCP- M1M00, AF3100OB, AF3140MB, AF3140MB-M0, AF3140MB-2, AF3140MB-M1, AF3140MB-PW, AF3140MB-P, AF3140MB-D12I, AF3140MBP-2, AF3140MC, AF3140MCP, AF3140MCPF, AF3100MB, AF3100MB-M0, AF3100MB-2, AF3100MB-M1, AF3100MB-X, AF3100MBP, AF3100MB-M01, AF3100MBP-M01, AF3100MBP-2, AF3100MBP-D12, AF3100MCP; 4. AirFlow Pediatric Resuscitator REF #s: AF2100OB, AF2100MB-3, AF2102OB, AF2100MB-D5, AF2140OB, AF2102MBP-2, AF2102MBPW-D4, AF2102MBP-M3, AF2140MB-2, AFW2142OB, AF2140MB-D4, AF2140MB-D43, AF2140MB-P2, AF2144MB, AFW2142MB-2, AF2142MB-D3, AF2242MBP, AF2142MBS-I, AF2142MBP- 224OT, AF2140MBPW- D23, AF2140MB-I, AF2140MB-K, AF2140MB-S, AF2140MBP-K, AF2100MB-PW, AF2100MB-MMC, AF2102MBP, AF2100MB- PWTS4, AF2140MCPF, AF2100MC, AF2102MC, AF2142MC-M01, AF2140MC, AF2142MC-D34, AF2140MCP-D4, AF2140MC-PT, AF2140MCP-K, AF2142MCP, AF2100MCP, AF2102MB, AF2102MCP, AF2102MC-PT, AF2142OB, AF2142OBNS, AF2142MB, AF2142MB-3, AF2142MB-2, AF2142MB-M5E, AF2142MB-PW, AF2142MBP, AF2142MB-K, AF2142MBP-2, AF2142MC-3, AF2142MCP-D4, AF2140MB, AF2140MB-PW, AF2140MBP, AF1142OB, AF2100MB; 5. AirFlow Small Adult Resuscitator REF #s: AF5140MC-D6, AF5140MB-M5, AF5100MB-D5, AF5142MBP-M5, AF5140MBPW-D5, AF5140MBPWTD5, AF5140MBP-D5, AF5140MBP-D6, AF5140MBPW- M5, AF5140MBP-M5, AFN5140MB, AF5140MB-OT, AF5140MB-T, AFD5140MB-T, AF5140MB-K, AF5140MBS-R, AF5140MBS-P, AF5140MB-PT, AF5140MB-ST, AF5140MCP-K, AF5100MB-PW, AF5102MB-N, AF5140MBS-M5. 6. Ventlab, LLC Resuscitation Kit REF # MC1000.
Reason
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Identifiers
| code_info | UDI-DI: 10889483075814 (Item # AF1000MBN-20, Case UDI: 30889483075818); UDI-DI: 10889483076606 (Item # AF1100MB, Ca…UDI-DI: 10889483075814 (Item # AF1000MBN-20, Case UDI: 30889483075818); UDI-DI: 10889483076606 (Item # AF1100MB, Case UDI: 30889483076600); UDI-DI: 10889483076613 (Item # AF1100MB-MMC, Case UDI: 30889483076617); UDI-DI: 10889483075838 (Item # AF1100MB-N, Case UDI: 30889483075832); UDI-DI: 10889483088074 (Item # AF1100MBP, Case UDI: 30889483088078); UDI-DI: 10889483086209 (Item # AF1100MB-PWG1, Case UDI: 30889483086203); UDI-DI: 10889483086216 (Item # AF1100MBS, Case UDI: 30889483086210); UDI-DI: 10889483094365 (Item # AF1100MC, Case UDI: 30889483094369); UDI-DI: 10889483159217 (Item # AF1100MC-P, Case UDI: 30889483159211); UDI-DI: 10889483071809 (Item # AF1140MB, Case UDI: 30889483071803); UDI-DI: 10889483230626 (Item # AF1140MBC-N Case UDI: 30889483230620); UDI-DI: 10889483090718 (Item # AF1140MB-D5, Case UDI: 30889483090712); UDI-DI: 10889483106372 (Item # AF1140MB-E, Case UDI: 30889483106376); UDI-DI: 10889483095096 (Item # AF1140MB-I, Case UDI |
|---|