FDA Device recall Z-0360-2016
Perseus A500 Anesthesia Workstation, MK06000
- FDA Device
- Class I
- terminated
- Published 2015-12-23
On 2015-12-23 the FDA classified a Class I recall of Perseus A500 Anesthesia Workstation, MK06000 by Draeger Medical, citing there is a potential problem with the device main power switch of the Perseus A500 Anesthesia Workstation. The device power switch may spontaneously fail during use, causing one o….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0360-2016 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2015-11-12 |
| Published | 2015-12-23 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Draeger Medical |
| Distribution | IL, ME, NY, OH, PA, WI |
Product
Perseus A500 Anesthesia Workstation, MK06000. Intended for use in anesthetizing adults, children, and neonates and can be used for automatic and manual ventilation, pressure-supported spontaneous breathing, and spontaneous breathing.
Reason
There is a potential problem with the device main power switch of the Perseus A500 Anesthesia Workstation. The device power switch may spontaneously fail during use, causing one of the following to occur: 1. The workstation has a fresh gas delivery failure and a ventilator failure with respective alarms. 2. The workstation shuts down completely with an audible alarm.
Identifiers
| code_info | ASHB-0140 ASHF-0039 ASHF-0055 ASHF-0056 ASHF-0057 ASHH-0008 ASHH-0009 ASHH-0010 ASHH-0011 ASHH-0012 ASHH-0013…ASHB-0140 ASHF-0039 ASHF-0055 ASHF-0056 ASHF-0057 ASHH-0008 ASHH-0009 ASHH-0010 ASHH-0011 ASHH-0012 ASHH-0013 ASHH-0014 ASHJ-0095 ASHJ-0096 ASHJ-0097 ASHJ-0098 ASHJ-0099 ASHK-0001 ASHK-0002 ASHK-0003 ASHK-0008 ASHK-0009 ASHK-0010 ASHK-0056 ASHK-0057 ASHK-0079 ASHK-0080 ASHK-0081 ASHK-0082 ASHK-oo83 ASHK-0084 ASHK-0085 ASHK-0086, and ASHK-0087. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.