FDA Device recall Z-0357-2024
GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004, 00024345004
- FDA Device
- Class II
- ongoing
- Published 2023-11-29
On 2023-11-29 the FDA classified a Class II recall of GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004, 00024345004 by Instrumentation Laboratory, citing internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 m….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0357-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-09-29 |
| Published | 2023-11-29 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Instrumentation Laboratory |
| Distribution | MA, MN, NY |
Product
GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004, 00024345004
Reason
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
Identifiers
| code_info | Part No. 00024307504: UDI-DI: 08426950087922, Lot Numbers: 324230M, 323630G; Part No. 00024330004: UDI-DI: 08426950082…Part No. 00024307504: UDI-DI: 08426950087922, Lot Numbers: 324230M, 323630G; Part No. 00024330004: UDI-DI: 08426950082507, Lot Numbers: 325655C, 322355C, 323755H, 323855E, 326255D; Part No. 00024345004: UDI-DI: 08426950082538, Lot Numbers: 325660C, 324160J, 323760H |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0357-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.