CAS I/II Absorbers that are parts of the BleaseSirius Anesthesia Workstation, the BleaseFocus Anesthesia Workstation with the following par…

On 2013-11-27 the FDA classified a Class I recall of CAS I/II Absorbers that are parts of the BleaseSirius Anesthesia Workstation, the BleaseFocus Anesthesia Workstation with the following par… by Del Mar Reynolds Medical, citing spacelabs Healthcare Ltd. of United Kingdom recalls , and Service Kits, Part Number 050-0659-00 and 050-0901-00. The Bag to-Vent switch in CAS I/II Absorbers may fail due to loose….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0352-2014
ClassificationClass I
Statusterminated
Initiated2013-10-17
Published2013-11-27
Last updated2026-09-13 11:53 UTC
CompanyDel Mar Reynolds Medical
DistributionUS

Product

CAS I/II Absorbers that are parts of the BleaseSirius Anesthesia Workstation, the BleaseFocus Anesthesia Workstation with the following part numbers: 12200900; 12200901; 12200902; 12200903; 14200100, and service kits part number 050-0659-00 and 050-0901-00. BleaseFocus Anesthesia Workstation: The Blease Frontline Plus Range, Anesthesia Machines are intended for use in the hospital environment and locations not requiring portability. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of a mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. This device is intended for use only by a suitably qualified physician. BleaseSirius Anesthesia Workstation: The Spacelabs BleaseSirius Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician.

Reason

Spacelabs Healthcare Ltd. of United Kingdom recalls , and Service Kits, Part Number 050-0659-00 and 050-0901-00. The Bag to-Vent switch in CAS I/II Absorbers may fail due to loose fastening hardware in an absorber.

Identifiers

code_info
SERIAL NUMBERS IN THE US: ABSO-001754, ABSO-004081, ABSO-100111, ABSO-100112, ABSO-100113, ABSO-100114, ABSO-100117,…SERIAL NUMBERS IN THE US: ABSO-001754, ABSO-004081, ABSO-100111, ABSO-100112, ABSO-100113, ABSO-100114, ABSO-100117, ABSO-100120, ABSO-100121, ABSO-100122, ABSO-100123, ABSO-100124, ABSO-100125, ABSO-100126, ABSO-100127, ABSO-100128, ABSO-100132, ABSO-100134, and ABSO-100135. SERIAL NUMBERS INTERNATIONAL: ABSO-100098, ABSO-100099, abso-100100, ABSO-100101, abso-100102, ABSO-100103, abso-100104, abso-100105, abso-100106, ABSO-100107, ABSO-100108, ABSO-100109, ABSO-100110, ABSO-100129, ABSO-100130, ABSO-100131, ABSO-100136, AND ABSO-100137.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0352-2014%22