FDA Device recall Z-0351-2025

Brand Name: Olympus Product Name: Cystoscope Outer Sheath Model/Catalog Number: WA22810A Software Version: N/A Product Description: Outer s…

On 2024-11-20 the FDA classified a Class II recall of Brand Name: Olympus Product Name: Cystoscope Outer Sheath Model/Catalog Number: WA22810A Software Version: N/A Product Description: Outer s… by Olympus, citing olympus is removing the statement of compatibility with a GreenLight Laser for BPH therapy from the Instruction for Use (IFU) due to the potential for a damaged tip during use of….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0351-2025
ClassificationClass II
Statusongoing
Initiated2024-09-13
Published2024-11-20
Last updated2026-09-13 11:57 UTC
CompanyOlympus
DistributionUS

Product

Brand Name: Olympus Product Name: Cystoscope Outer Sheath Model/Catalog Number: WA22810A Software Version: N/A Product Description: Outer sheaths can be combined with rigid Telescopes. Component: N/A

Reason

Olympus is removing the statement of compatibility with a GreenLight Laser for BPH therapy from the Instruction for Use (IFU) due to the potential for a damaged tip during use of a laser probe.

Identifiers

code_infoModel: WA2280A, UDI: 04042761051729, Lot#: All lots.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0351-2025%22