Puritan Bennett 980 Series Ventilator, 980X1ENDIUU
- FDA Device
- Class I
- terminated
- Published 2021-12-15
On 2021-12-15 the FDA classified a Class I recall of Puritan Bennett 980 Series Ventilator, 980X1ENDIUU by Covidien, citing manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0347-2022 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2021-11-04 |
| Published | 2021-12-15 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Covidien |
| Distribution | US |
Product
Puritan Bennett 980 Series Ventilator, 980X1ENDIUU
Reason
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Identifiers
| code_info | GTIN/UDI - 10884521171541 Serial Numbers: 35B1700300 35B1700306 35B1700307 35B1700362 35B1700372 35B1700384 35B1700387…GTIN/UDI - 10884521171541 Serial Numbers: 35B1700300 35B1700306 35B1700307 35B1700362 35B1700372 35B1700384 35B1700387 35B1700388 35B1700396 35B1700397 35B1700403 35B1700413 35B1700416 35B1700418 35B1700426 35B1700431 35B1700434 35B1700440 35B1700445 35B1700451 35B1700452 35B1700453 35B1700459 35B1700461 35B1700464 35B1700468 35B1700472 35B1700477 35B1700478 35B1700479 35B1700480 35B1700484 35B1700485 35B1700494 35B1700495 35B1700496 35B1700501 35B1700502 35B1700503 35B1700504 35B1700534 35B1700574 35B1700579 35B1700604 35B1700609 35B1700638 35B1700655 35B1700664 35B1700816 35B1700919 35B1700994 35B1701001 35B1701418 35B1701698 35B1702832 35B1703438 35B1800033 35B1800721 35B1800811 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Other recalls by Covidien
- EndoFLIP System that includes: ENDOFLIP 1 SYSTEM EF-100 05391530810050 10884521809307 REFURBISHED SYSTEM RFG-EF100 ENDOFLIP 1 1088452180809…
FDA Device · Class II · 2026-08-26 · Covidien - V-Loc 90 Absorbable Wound Closure Device VLOC 90 0 VL 22 GS-21 VLOCM0336 VLOC 90 0 VL 22 GS-25 VLOCM0436 VLOCM0436 VLOCM0003 V-LOC* 90 4-0…
FDA Device · Class II · 2026-08-19 · Covidien - V-Loc 180 Absorbable Wound Closure Device BOX BEDSIDE STAPLING IS KING BOX06614V4 BOX BY PASS FINAL BOX07405V1 BOX BYPASS CARDON BOX01056V2…
FDA Device · Class II · 2026-08-19 · Covidien - Dermalon 88861757-31 DERMALON* 4-0 BLU 75CM C13 88861757-31 88861799-31 DERMALON* 4-0 BLU 75CM C14 88861799-31
FDA Device · Class II · 2026-08-19 · Covidien - Surgipro II S8698GX SURGIPRO II BL 5/0 18 P-13 S8698GX
FDA Device · Class II · 2026-08-19 · Covidien
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0347-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.