FDA Device recall Z-0346-2024
Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Mo…
- FDA Device
- Class II
- ongoing
- Published 2023-11-29
On 2023-11-29 the FDA classified a Class II recall of Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Mo… by Aizu Olympus, citing channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the end….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0346-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-09-25 |
| Published | 2023-11-29 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Aizu Olympus |
| Distribution | US |
Product
Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Model Numbers: TJF-145, TJF-150, TJF-Q180V, TJF-Q190V.
Reason
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Identifiers
| code_info | Model Number (UDI-DI): TJF-145 (04953170339950), TJF-150 (04953170307553), TJF-Q180V (04953170229503), TJF-Q190V (04953170405563). |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0346-2024%22