RecallRadar

FDA Device recall Z-0346-2024

Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Mo…

On 2023-11-29 the FDA classified a Class II recall of Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Mo… by Aizu Olympus, citing channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the end….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0346-2024
ClassificationClass II
Statusongoing
Initiated2023-09-25
Published2023-11-29
Last updated2026-09-13 11:56 UTC
CompanyAizu Olympus
DistributionUS

Product

Duodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Model Numbers: TJF-145, TJF-150, TJF-Q180V, TJF-Q190V.

Reason

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Identifiers

code_infoModel Number (UDI-DI): TJF-145 (04953170339950), TJF-150 (04953170307553), TJF-Q180V (04953170229503), TJF-Q190V (04953170405563).

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0346-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.