Transnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach

On 2023-11-29 the FDA classified a Class II recall of Transnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach by Aizu Olympus, citing channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the end….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0344-2024
ClassificationClass II
Statusongoing
Initiated2023-09-25
Published2023-11-29
Last updated2026-09-13 11:56 UTC
CompanyAizu Olympus
DistributionUS

Product

Transnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach. Model Number: PEF-V.

Reason

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Identifiers

code_infoModel Number (UDI-DI): PEF-V (04953170340376).