RecallRadar

FDA Device recall Z-0341-2024

Colonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve)

On 2023-11-29 the FDA classified a Class II recall of Colonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve) by Aizu Olympus, citing channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the end….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0341-2024
ClassificationClass II
Statusongoing
Initiated2023-09-25
Published2023-11-29
Last updated2026-09-13 11:56 UTC
CompanyAizu Olympus
DistributionUS

Product

Colonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve). Model Numbers: PCF-H180AI, PCF-H180AL, PCF-H190DI, PCF-H190DL, PCF-H190I, PCF-H190L, PCF-H190TI, PCF-H190TL, PCF-HQ190I, PCF-HQ190L, PCF-PH190I, PCF-PH190L, PCF-Q180AI, PCF-Q180AL, PCF-S.

Reason

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Identifiers

code_info
Model Number (UDI-DI): PCF-H180AI (04953170339875), PCF-H180AL (04953170339899), PCF-H190DI (04953170363696), PCF-H190D…Model Number (UDI-DI): PCF-H180AI (04953170339875), PCF-H180AL (04953170339899), PCF-H190DI (04953170363696), PCF-H190DL (04953170363672), PCF-H190I (04953170305214), PCF-H190L (04953170305191), PCF-H190TI (04953170420719), PCF-H190TL (04953170420702), PCF-HQ190I (04953170416132), PCF-HQ190L (04953170416118), PCF-PH190I (04953170305252), PCF-PH190L (04953170305238), PCF-Q180AI (04953170307713), PCF-Q180AL (04953170307720), PCF-S (04953170339912).

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0341-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.