RecallRadar

FDA Device recall Z-0336-2022

Cardiosave Hybrid IntraAortic Balloon Pump

On 2021-12-22 the FDA classified a Class I recall of Cardiosave Hybrid IntraAortic Balloon Pump by Datascope, citing firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which c….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0336-2022
ClassificationClass I
Statusongoing
Initiated2021-11-15
Published2021-12-22
Last updated2026-09-13 11:56 UTC
CompanyDatascope
DistributionUS

Product

Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-UC-0800-34, 0998-00-0800-35, 0998-UC-0800-35, 0998-00-0800-36, 0998-UC-0800-36, 0998-00-0800-45, 0998-UC-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65, 0998-UC-0800-65.

Reason

Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.

Identifiers

code_info
All Serial Numbers. May also be designated as part number D998 etc. Model Number, UDI-DI: 0998-00-0800-31, 106075671…All Serial Numbers. May also be designated as part number D998 etc. Model Number, UDI-DI: 0998-00-0800-31, 10607567109053; 0998-UC-0800-31, 10607567109053; 0998-00-0800-32, 10607567111117; 0998-UC-0800-32, 10607567111117; 0998-00-0800-33, 10607567109008; 0998-UC-0800-33, 10607567109008; 0998-00-0800-34, 10607567111940; 0998-UC-0800-34, 10607567111940; 0998-00-0800-35, 10607567109107; 0998-UC-0800-35, 10607567109107; 0998-00-0800-36, 10607567114187; 0998-UC-0800-36, 10607567114187; 0998-00-0800-45, 10607567108421; 0998-UC-0800-45, 10607567108421; 0998-00-0800-52, 10607567108438; 0998-UC-0800-52, 10607567108438; 0998-00-0800-53, 10607567108391; 0998-UC-0800-53, 10607567108391; 0998-00-0800-55, 10607567108414; 0998-UC-0800-55, 10607567108414; 0998-00-0800-65, 10607567113432; 0998-UC-0800-65, 10607567113432

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0336-2022%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.