FDA Device recall Z-0336-2014
Vascutek Gelsoft Vascular Graft (peripheral) Labeling reads in part, "***Gelsoft Gelatin Impregnated Knitted Vascular Prosthesis***STRAIGHT…
- FDA Device
- Class II
- terminated
- Published 2013-11-27
On 2013-11-27 the FDA classified a Class II recall of Vascutek Gelsoft Vascular Graft (peripheral) Labeling reads in part, "***Gelsoft Gelatin Impregnated Knitted Vascular Prosthesis***STRAIGHT… by Terumo Cardiovascular Systems, citing terumo Cardiovascular Systems (Terumo CVS) became aware of a customer complaint that a 15mm x 8mm Gelsoft graft was received which the customer believed was a 15mm x 6mm Gelsoft g….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0336-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-10-23 |
| Published | 2013-11-27 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | CA, NC |
Product
Vascutek Gelsoft Vascular Graft (peripheral) Labeling reads in part, "***Gelsoft Gelatin Impregnated Knitted Vascular Prosthesis***STRAIGHT***STERILE***EO***Item No: 631508***Diameter 8mm***Usable Length 15cm***Manufactured By***VASCUTEK Ltd. a TERUMO Company Newmains Avenue Inchinnan, Renfrewshire PA4 9RR Scotland***" The Vascutek Gelsoft Vascular Prosthesis is a gelatin-sealed, woven polyester graft indicated for abdominal and peripheral vascular repair, i.e. replacement or bypass in aneurysmal and occlusive disease of arteries
Reason
Terumo Cardiovascular Systems (Terumo CVS) became aware of a customer complaint that a 15mm x 8mm Gelsoft graft was received which the customer believed was a 15mm x 6mm Gelsoft graft unit. The manufacturer confirmed a batch of product code 631508 was incorrectly labeled.
Identifiers
| code_info | Catalog Number 631508 Serial Numbers: 0001765391, 0001765392, 0001765393, 0001765394, 0001765395, 0001765396, 0001765397, 0001765398, 0001765399 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0336-2014%22