RecallRadar

FDA Device recall Z-0330-2018

CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system

On 2018-01-17 the FDA classified a Class II recall of CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system by Philips Electronics North America, citing the tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0330-2018
ClassificationClass II
Statusterminated
Initiated2017-10-19
Published2018-01-17
Last updated2026-09-13 11:54 UTC
CompanyPhilips Electronics North America
Distributionn/a

Product

CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography

Reason

The tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in specified distance.

Identifiers

code_infoAll CombiDiagnost R90 UDI 709030

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0330-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.