FDA Device recall Z-0329-2025
SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198
- FDA Device
- Class II
- ongoing
- Published 2024-11-20
On 2024-11-20 the FDA classified a Class II recall of SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198 by Wissner Bosserhoff, citing when the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0329-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-07-11 |
| Published | 2024-11-20 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Wissner Bosserhoff |
| Distribution | US |
Product
SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198
Reason
When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.
Identifiers
| code_info | UDI/DI 4050621100401, 19-0990 UDI/DI 4050621100012, 19-1062 UDI/DI 4050621100432, 19-1152 UDI/DI 4050621100449, 19-1119…UDI/DI 4050621100401, 19-0990 UDI/DI 4050621100012, 19-1062 UDI/DI 4050621100432, 19-1152 UDI/DI 4050621100449, 19-1119 UDI/DI 4050621100470, 19-1155 UDI/DI 4050621100487, 19-1156 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0329-2025%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.