RecallRadar

FDA Device recall Z-0328-2024

Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures wit…

On 2023-11-29 the FDA classified a Class II recall of Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures wit… by Aizu Olympus, citing channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the end….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0328-2024
ClassificationClass II
Statusongoing
Initiated2023-09-25
Published2023-11-29
Last updated2026-09-13 11:56 UTC
CompanyAizu Olympus
DistributionUS

Product

Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, esophagus and surrounding organs Model Numbers: BF-UC180F, BF-UC190F.

Reason

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Identifiers

code_infoModel Number (UDI-DI): BF-UC180F (04953170356360), BF-UC190F (04953170399831).

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0328-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.