FDA Device recall Z-0326-2024

Alinity c Iron Reagent, Reference Number 08P3920

On 2023-11-29 the FDA classified a Class II recall of Alinity c Iron Reagent, Reference Number 08P3920 by Sentinel Ch Spa, citing A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0326-2024
ClassificationClass II
Statusongoing
Initiated2023-09-29
Published2023-11-29
Last updated2026-09-13 11:56 UTC
CompanySentinel Ch Spa
DistributionUS

Product

Alinity c Iron Reagent, Reference Number 08P3920

Reason

A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.

Identifiers

code_infoUDI-DI: 00380740136253; Lot Number: 21046Y600, 21118Y600, 30227Y600, 30302Y600

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.