FDA Device recall Z-0319-2017
IMMULITE/ IMMULITE 1000 Systems CMM CMV IgM Assay, REF/Catalog Number LKCM1(D) - 100 tests (US), Siemens Material Number (SMN) 10370301 (US…
- FDA Device
- Class II
- terminated
- Published 2016-11-02
On 2016-11-02 the FDA classified a Class II recall of IMMULITE/ IMMULITE 1000 Systems CMM CMV IgM Assay, REF/Catalog Number LKCM1(D) - 100 tests (US), Siemens Material Number (SMN) 10370301 (US… by Siemens Healthcare Diagnostics, citing multiple customers reported seeing an increase in imprecision with some patients resulting in high % CVs with the IMMULITE 2000/2000XPi CMV (IgM) assay starting with reagent lots….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0319-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-08-30 |
| Published | 2016-11-02 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
IMMULITE/ IMMULITE 1000 Systems CMM CMV IgM Assay, REF/Catalog Number LKCM1(D) - 100 tests (US), Siemens Material Number (SMN) 10370301 (US), Unique Device Identification Number (UDI) 00630414985718 (US); and REF/Catalog Number LKCM1 - 100 tests (OUS); SMN 10381296 (OUS), UDI 00630414963747 (OUS); IVD --- Made in UK Siemens Healthcare Diagnostics Products Ltd., Glyn Rhonwy, Llanberis, Gwynedd, LL55 4EL UK For in vitro diagnostic use with the IMMULITE Systems Analyzers for the qualitative detection of IgM antibodies to Cytomegalovirus (CMV) in human serum or plasma (EDTA or heparinized), as an aid in the determination of an acute CMV infection.
Reason
Multiple customers reported seeing an increase in imprecision with some patients resulting in high % CVs with the IMMULITE 2000/2000XPi CMV (IgM) assay starting with reagent lots 259 and above.
Identifiers
| code_info | UDI 00630414985718/Catalog # LKCM1(D): Kit lot D0330; UDI 00630414963747/Catalog # LKCM1: Kit lots 0330, 0331 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.